ICP-OES Metals Testing: What It Measures and When to Use It

H
Helen Micklewright
Founder
2026-01-11
lab, analysis, diagnostics, hospital, test tube, test, tests, medical, the doctor, test tubes, medic, treatment, patient

ICP-OES (inductively coupled plasma optical emission spectrometry) is the standard method for quantifying metals in liquid samples. It is used in environmental monitoring, food safety, cosmetic ingredient compliance, and industrial quality control. This article explains how it works, what preparation your sample needs, and what the result actually tells you.

How the instrument works

The sample is introduced as a fine mist into a plasma torch burning at around 6000-8000 K. At that temperature, atoms in the sample are excited and emit light at characteristic wavelengths. The instrument measures the intensity of light at each wavelength, and the intensity is proportional to the concentration of the corresponding element.

The method can measure up to 24 elements simultaneously in a single run, which makes it efficient for multi-element panels. Detection limits are typically in the low microgram-per-litre range for most elements.

Sample preparation

Most solid samples need to be digested into solution before analysis. This involves treating the sample with concentrated acid (typically nitric acid, sometimes with hydrofluoric acid for silicate matrices) under heat or microwave conditions. The digestion method affects which elements are recovered and at what efficiency.

Liquid samples may be analysed directly or after dilution, depending on the expected concentration range. If the sample contains organic solvents, these need to be removed or the method adapted, as high organic content can affect the plasma stability.

Common applications

In cosmetics, ICP-OES is used to check heavy metal content (lead, arsenic, cadmium, mercury) against EU Cosmetics Regulation limits. In food, it is used for nutritional labelling of minerals and for contaminant monitoring.

In environmental work, it is the standard method for metals in soil and water under UK Environment Agency guidance. In industrial quality control, it is used to verify the composition of alloys, plating solutions, and process chemicals.

What the result tells you and what it does not

ICP-OES gives you the total concentration of each element in the digested sample. It does not tell you which compound the element is in (speciation), or whether the element is bioavailable. For speciation (for example, distinguishing hexavalent chromium from trivalent chromium), a different method is needed.

The result is only as good as the digestion. If the digestion did not dissolve all the sample, the result will be lower than the true value. A good report will state the digestion method and the certified reference material used to verify recovery.

If you are not sure whether ICP-OES is the right method for your metals question, a short description of the sample matrix and the elements of concern is enough for Helen to confirm the approach.

#ICP-OES#Metals testing#Environmental analysis#Cosmetic compliance

Related reading

See all news

Precision work. Quiet results.

Page·→
Home Home Learn more about what we do. Read more
About The story behind the lab Meet the people behind the work. Read more
Contact How to reach us Come visit, or drop us a line. Read more
Privacy Privacy Learn more about what we do. Read more
Terms Terms Learn more about what we do. Read more